
MAJOR ASSEMBLIES
Download PDF
SYSTEM FEATURES
ASSEMBLY FEATURES
DOCUMENTATION
OPTIONS
SAMPLE CONDITIONING SKID
The Sample Conditioning Skid is located down-stream from the Isokinetic Sample Probe Assembly. Each of the two sample flow paths is provided with two prefilters and one sample filter. The selectable prefilters prevent contamination of the detectors on the Sample Detection Skid by filtering particulates and iodine. The sample filter provides particulates and iodine sample for laboratory analysis. Solenoid valves control the sample flow path by operator request. This configuration allows filters to be changed quickly without affecting monitor operation (all filters have quick-disconnect fittings). Filters on the mid/high-range flow path have 4π solid cast-lead shielding enclosure to minimize exposure to personnel (see Flow Diagram). A sample can be collected or a prefilter path selected locally with switches in the electrical enclosure or remotely with switches in the Control room Display Assembly. The Sample Conditioning Skid should be floor-mounted as closely as possible to the sample point to minimize plate-out and provide easy access for sample retrieval.
SAMPLE DETECTION SKID
The Sample Detection Skid is located down-stream from the Sample conditioning Skid. It monitor radioactive gas concentrations over a 12-decade range using three radiation detectors. Each of the two flow paths has separate pumps, valves flowmeters, relays, and control switches. The flow through the monitor is automatically controlled by the RM-80 Microprocessor to maintain isokinetic sample flow rates from the Isokinetic Sample Probe Assembly. Each detector has a radio-active checksource to verify proper operation, and a 6in. lead shield to reduce background radiation effects. Each detector has a hinged, lead door to allow easy access to the sample canister or the detector. The low concentration flow path uses the RD-52 Gas Detector which contains a thin plastic scintilla-ion detector. The mid/high concentration flow path uses the RD-72 Dual-Range Gas Detector which has two beta/gamma-sensitive solid-state CdTe detectors (one each for mid-and hig-range detection). Range overlap provides continuity of measurement. A Customer interface Junction Box (CIJB) houses alarm relays and terminal blocks. detector signal and communications wiring is in another junction box to minimize electrical noise. Active electronics including power supplies and preamplifiers are located in a lead shielding enclosure to ensure their survival through high radiation doses. the Power Control Center accepts incoming 120 V ac power, distributes power to the Sample Detection Skid, and provides a main power disconnect switch. The Sample Detection Skid can be floor-mounted next to the Sample Conditioning Skid or up to 50 away.
![]() |
![]() |
Xe133 Concentration (_Ci/cm3) |
Low-Range Pump Status |
Mid/High-Range Pump Status |
Purge |
Detector Data Used by RM-80 Microprocessor |
Increasing |
|
|
|
|
10-7 to 5 x 10-2 |
On |
Off |
- |
Low-range |
5 x 10-2 to 10-1 |
On |
On |
- |
Low-range |
10-1 to 10 |
On |
On |
Low-range purged at 10 _Ci/cm3 |
Mid-range |
10 to 102 |
Off |
On |
- |
Mid-range |
102 to 105 |
Off |
On |
- |
High-range |
|
|
|
|
|
Decreasing |
|
|
|
|
102 to 10 |
Off |
On |
- |
Mid-range |
10 to 10-1 |
On |
On |
- |
Mid-range |
10-1 to 5 x 10-2 |
On |
On |
Mid/High-range purged at 5 x 10-2 _Ci/cm3 |
|
10-2 to 10-7 |
On |
Off |
- |
Low-range |
OPTICAL ISOKINETIC SAMPLE PROBE ASSEMBLY
The optional isokinetic Sample Probe Assembly obtains representative sample from the vent. Accordingly, the WRGM has two sets of probes, one set for each of the two sample flow paths. Custom isokinetic probes are usually necessary for compatibility with different installations. Sorrento Electronics has extensive experience in custom isokinetic probe designs. Designs are based on effluent point parameters such as flow rate, cross-section, construction material, access, and temperature. The probe design balances technical refinement with reasonable cost. Typically, all probes, including a vent flow sensor for automatic isokinetic flow control, are mounted on a support structure that adapts to existing installation, such as flanges or plates.
OPTIONAL MULTIPOINT SAMPLE SELECTOR
An optional Multipoint Sample Selector (adjacent photo) allows a WRGM to monitor up to three effluent paths. The effluent path to be monitor is selected at the Control Room Display Assembly.
SERVICING
The WRGM is easy to service. Prefilters and samples can be replaced in less than 5 minutes without special tools. High activity sample can be transported to the radiochemistry laboratory for analysis inside the filter‘s 4π solid lead shielding, minimizing radiation exposure to personnel retrieving the samples. isolation valves isolate selected parts of the monitor for service.
A portable RM-23P Control/display module, included with the WRGM, is used to help maintain and calibrate digital radiation monitors (see adjacent photo). The RM-23P has the same capabilities as the RM-23A Control/Display Module in the Control Room Display Assembly. The RM-23P can be connected locally to the Sample Detection Skid or the RM-80 Microprocessor/CIJB.
![]() |
![]() |
RM-80 MICROPROCESSOR/CUSTOMER INTERFACE JUNCTION BOX
The RM-80 Microprocessor performs data acquisition and control functions, and provides alarm relays, analog outputs, and digital communications. The RM-80 Microprocessor data base, which can be programmed for specific applications by the user, contains the calibration constants for the RD-52 Gas Detector and RD-72 Dual-Range Gas Detector, alarm setpoints, and history files (twenty-four 10-min averages, twenty-four 1-hr averages, and twenty-eight 24-hr averages). This allows the monitor to be customized to its particular application. The RM-80 Microprocessor/CIJB can be located up to 500 ft from the Sample Detection Skid. The CIJB contains terminal blocks for all field cables, which enter through top entry conduit fittings.
CONTROL ROOM DISPLAY ASSEMBLY
The Control Room Display Assembly provides remote control and display for the system. It is contained in a 19-in. wide modified NIM bin that fits in a standard control room rack. The Control Room Display Assembly consists of a micro-processor-based, RM-23A Control/Display Module and an operator control panel. The RM-23A Control/Display Module continuously polls the RM-80 Microprocessor for the radiation level and operation status. All relevant monitor parameters, such as gas concentration (_Ci/ cm3), effluent release rate (_Ci/sec), alarm status, flow status, purge actuation, calibration constants, and historical averages, can be displayed on the RM-23A Control/Display Module. Only one of four channels can be displayed on the RM-23A Control/Display Module at one time. Channels one through three are the gas concentration of the low-, mid-, and high-range detectors in _Ci/cm3. Channel four is the effluent release rate of the active detector channel display in _Ci/sec. The operator control panel is next to the RM-23A Control/Display Module allowing the operator to control the prefilters and samplers on the Sample Conditioning Skid. Control switches allow the operator to change a flow path through the prefilters as necessary. Sample timers allow the operator to select sampling times during routine or accident conditions. To obtain a sample, the operator sets the sample timer and presses the actuation push-button; the timed sample is then taken automatically. A key switch is provided for supervisory control of the RM-23A control/Display Module. Customer cable connections are made through terminal blocks on the rear of the bin. One strip-chard recorder gives a hard copy of the three radioactive gas concentration levels. A second strip-chard recorder gives a hard copy of the effluent release rate.
![]() |
![]() |
SPECIFICATIONS SYSTEM Operational Range: Pressure - Atmospheric; Temperature
-35° to 122°F (1.7° to 55°C); Humidity - to 100% (noncondensing)
Range: 10-7 to 105 _Ci/cm3 (Xe133) with 1 mr/hr background
Sample Temperature: 120°F (49°C) maximum at 90% RH; 130°F (54°C) maximum at 70% RH
Sample Pressure: 15psig (1.05 kgcm2) maximum
Sample Flow: Low-range - 1.7 ft3/min (800 cm3/sec) nominal; mid/high-range - 0.06 ft3/min (28 cm3/sec) nominal
Sample Connections: Low-range - .5-in. tube fitting; mid/high-range - .25-in. tube fitting; common return line - .5in. tube fitting; connecting lines between Isokinetic Sample Probe Assembly manifold and Sample conditioning Skid by customer per ANSI N13,1-1969.
Power: 120 V ac, 50 to 60 Hz, 15 A nominal at the main power disconnect switch on the Sample Detection Skid which provides power to the Sample Detection Skid and Sample conditioning Skid; RM-80 Microprocessor/CIJB, 1 A nominal; Control room Display Assembly, 3 A nominal
Overall System Accuracy: Meets or exceed U.S. Regulatory Guide 1.97 (Rev. 3)
Quality Assurance: Meets 10 CFR 50, Appendix B; ANSI N45.2; Canadian Standard Z299.3
SAMPLE CONDITIONING SKID
Filter Assemblies: Low-range - 5-in. inline quick-disconnect; mid/high-range - .25-in. inline quick-disconnect
Filters: 1.9-in. diameter HV LB 5211 fiberglass (Particulate);
silver zeolite cartridge (Iodine); replacement time - less than 5 min
Dimensions: 43.5 in. high x 48 in. wide x 30 in. deep
Weight: 1600 lb (730 kg) approximate
Plumbing: Type 304 stainless steel for all wetted parts except gaskets
Shielding: 2 in. lead shielding with mid/high-range filters to minimize personnel exposure
Integrated Dose: 105 Rad
SAMPLE DETECTION SKID
Detectors: 2-in.-diameter x 0.01-in.-thick beta phosphor; 2in.-diameter photomultiplier tube (low range); two CdTe solid-state, beta-gamma detectors with integral preamplifiers (mid and high range)
| About Us | News & Information | Contact | Support | Procurement | Quality | Careers Energy Products | Terminal Automation Products | Radation Monitoring Systems | TRIGA | Gulftronic |
©General Atomics, All Rights Reserved
For technical issues regarding this website contact webmaster@ga.com